Product Name:Dental Ray X System. Working principle:The high voltage generator provides high voltage to both ends of the Ray X.
Tube filament and metal target in the internal Ray X source assembly, a large number of electrons are. Generated on the cathode filament of the Ray X tube and move at a high speed in the vacuum tube. Impact the metal target to generate the Ray X. The Ray X penetrate the different tissue densities of the body, such as teeth and muscles after pass through the Ray X window and beam tube, the Ray X which. Carry image information through body tissue, are passed through image receiving devices such as. Phosphor sheets, film or digital image receivers, to show different densities of human tissue images. Application Range: used for the teeth to taking an Ray X photography, with an intraoral image.Receiver, images are obtained for the clinical diagnosis. Structural composition: the product consists of the host including the Ray X tube, controller and.
Rechargeable battery, beam limiting tube, power adapter, stand and exposure switch. Power Adapter Input Voltage Uniphase AC 100 ~240V, ±10% Supply Frequency 50/60Hz, ±1Hz Input Power 286W. Charge voltage DC 25.2V. Nominal value of the focal point 0.8mm. Available maximum tube voltage at 2mA 60kV.
Maximum output electric power 0.12kW. Maximum loading factor combination 60kV? Repeatability of radiation output No more than 0.5. Tube voltage (The deviation is within the range of ±10%) 60kV. Tube current (The deviation is within the range of ±20%) 2mA. Running mode Discontinuous operation, maximumexcitation (open) time : minimumexcitation (close) time = 1:30. Total filtration of Ray X tube assembly >2.5mm Al. Intrinsic filtration of Ray X tube assembly 0.75mm Al. Additional filtration 1.00mm Al. Adjustment range of the loading time 0.01 ~2.00s, adjustable. Distance of the Focal point to skin (The deviation is within 05%). Ray X field size at the end of beam limiting tube round , ?Nominal electric power 0.12kW(60kV? Available maximum tube current at 60kV 2mA. Inverter frequency of the high voltage generator 70kHzNote: the loading time adjustment range meets the R10 series for the adjustment. Or internal power supply class , does not work while charging.
(2) Classified by electric shock protection type:Type B application section. (3) Classified by the degree of protection against fluid intake:Common equipment. (5) Classified by the degree of safety for use with flammable anesthetic gases mixed with air or with oxygen or nitrous oxide:NonAP/APG type.
(6) Classified by operation mode:Discontinuous operation, Maximum excitation (open) time: Minimum excitation (close) time =1:30. (7) The product has no application part with protection against defibrillation discharge effect.
(8) Whether the equipment has a signal output or input part:No. (9) The product is a common equipment. (10) Permanent or non-permanent installation equipment:Non-permanent installation equipment.
(11) Radiation type:Ionizing radiation. (12) Radiation property:Ray X. (13) Radiation distribution:The Ray X radiates from the positive direction of 6cm circle in the window of the limited beam barrel when used. Ray X tube technical specification. The Ray X tube model XD10D-0.5 / 70.
Maximum working tube voltage 70kV. Anode angle of the Ray X tube 19º. Intrinsic filtration 0.75 mm Al. Filament current 2.0 A. Filament voltage 2.85 ± 0.5 V.
Food & Drug Administration and state and local regulatory agencies. If you have questions about legal obligations regarding sales of medical devices, you should consult with the FDA's Center for Devices and Radiological Health. This dental items is not meant to replace your doctor diagnostic. It is mainly for your piece of mind and comfort. Regulation Number: 21 CFR 872.6070. FDA for Handpiece : 510(K) Number: K181691, Regulation Number: 872.4200, Classification Product Code: EFB, Subsequent Product Code: EGS. FDA for Scaler : 510(K) Number: K163414, Regulation Number: 872.4850, Product Code: ELC. Company name: Dreamland Technology Ltd.